Send request

loading

Two-Channel Flashlamp Driver PS5053 for Pulsed Lasers

Flashlamp driver PS5053 is designed for flash-lamp pumped lasers and presents two channel device consisting of two capacitor charging, simmer/trigger and pulse forming modules and control circuit.

Features:

  • Output voltage up to 2000 V
  • Single unit for oscillator-amplifier systems
  • Built-in serial ignition circuit
  • Built-in simmer power supply
  • Internal/external triggering
  • Output voltage regulation ±0.1 %
  • RS232/CAN interface for remote control
  • Single phase mains
  • LabVIEW demo application

Flashlamp driver PS5053 is designed for flashlamp-pumped lasers and presents a two-channel device consisting of two capacitor charging, simmer/trigger and pulse forming modules and control circuit. It is suitable to be used in oscillatoramplifier laser systems.
This model is an updated version of flashlamp driver PS5012. PS5053 features microprocessor control and back illuminated LCD display where all output parameters of power supply are conveniently displayed. Driver can be remotely controlled through RS-232 and CAN (Controller Area Network) interface.

 

Number of independent outputs 2
Number of charging modules for first and second channels 1+1 1+2 1+3 or 2+2
Max. average output power Pavg at 10 Hz PRR, kJ/s 1), 2) 1.6 kJ/s 2.4 kJ/s 3.2 kJ/s
Standard charging voltage Uch 1000 V, 1400 V, 1800 V 3)
Pulse duration fixed
Pulse repetition rate < 150 Hz
Pulse to pulse stability 0.1 %
Load regulation 0.1 %
Linearity 0.2 %
Resolution 1 V
Ignition pulse voltage 16 kV4)
Ignition pulse duration > 1000 ns
Simmer current 0.6 A
Simmer voltage <300 V
Striking voltage <900 V
Protection features overvolt, overheat, flashlamp breakdown, interlock
Error report no simmer current, no charge, HV connectors
Remote control RS-232 / CAN (CANopen on request)
Standard C PFN value < 240 µF 1)
Mains Single phase 230 V (-10 %, + 6 %) or 3 phase 380 V (-10 %, +6%) 5)
Power consumption, average 3.2 kW 4.5 kW 5.8 kW
Power consumption, peak 4 kW 6 kW 8 kW
Operation conditions
Ambient temperature from 0 to + 40 °C
Humidity from 10 to 90 % non-condensing

1) Total for both channels
2) See chart in the next tab for other pulse repetition rates
3) Inquire for other voltages
4) Optional 30 kV
5) 3-phase 200 V or 208 V mains are optional

Specifications in table are given as reference. We always suggest to optimize power supply by customer's usage conditions

Ordering / part number information

Please indicate next points by inquiry:

  • Flash lamp type (bore diameter; gap length, gas type and pressure)
  • Maximal pulse energy
  • Pulse duration
  • Maximal pulse repetition rate